HS-480D type helium mass spectrometer leak detector

The HS-480D is a helium mass spectrometer leak detector with an oil-free dry pump, which is particularly suitable for leak detection of high-clean vacuum equipment, and it is designed in the form of a trolley, making it more convenient to move freely in the workplace. Selected high-quality imported vortex dry pump, capable of long-time continuous work. Our company promises to provide a warranty of up to 3 years for core components, allowing you to use with confidence.

Applications

Vacuum equipment manufacturing: ensure there are no leaks. Laboratory chemistry: check whether there are leaks in chemical reactors or vacuum dryers, ensuring experimental safety. Aerospace industry: check for leaks in spacecraft and rocket fuel systems and propulsion systems. Pharmaceutical industry: check for leaks in pharmaceutical equipment to ensure drug quality and safety. Electronics industry: check for leaks in vacuum tubes, semiconductor devices, display devices and other electronic components. Automotive manufacturing: check for leaks in car engines, brake systems, and air conditioning systems. Medical equipment: check for leaks in medical equipment (such as dialysis machines, respirators, etc.) to ensure the safety and effectiveness of medical equipment.

Features

--Selected high-quality imported oil-free dry pump, capable of long-time continuous work;
--Designed in the form of a trolley, convenient for moving.

  • Main technical indicators


    • Minimum detectable leak rate: 3×10-13Pa•m3/s

    • Leak rate display range: 1×10-2—1×10-13Pa•m3/s

    • Startup time: ≤3min

    • Response time: ≤1.0s

    • Maximum pressure at the leak detection port: 2000Pa

    • Limiting pressure: 5×10-4Pa

    • Mechanical pump suction speed: 0.5L/s

    • Power requirements: 220V, 50Hz, single phase, 10A

    • Working environment: 5-35℃

    • Relative humidity: ≤80%

    • External dimensions: 560(W)×420(D)×320(H)

    • Weight: 42kg


    Main configuration


    • Special molecular pump for German Pfeiffer leak detector

    • UK Edwards mechanical pump

    • Built-in standard leak hole

    • American AD amplifier

    • Using a mass spectrometry-specific module

    • Delta liquid crystal touch color screen